NIBP Validation Study
Non-Invasive Blood Pressure Validation Study
Philips Clinical & Medical Affairs Global
189 participants
Jun 23, 2023
OBSERVATIONAL
Conditions
Summary
The purpose of this study is to compare measurements of blood pressure (BP) between the Philips non-invasive blood pressure (NIBP) system (including NIBP cuff and portable patient monitor) and invasive radial for Subject Group 2 and 3, and radial or umbilical arterial line for neonates (A-line) in critical care patients.
Eligibility
Inclusion Criteria24
- Patient admitted into the NICU, PICU, ICU, or OR (operating room)
- Adult aged 18 years and older, or parent/legal guardian of minor, willing and able to understand and provide informed consent/assent
- Indication for NIBP cuff
- Indication for radial arterial line for Subject Group 2 and Subject Group 3, and radial or umbilical arterial line for Subject Group 1
- Meet the following stratification criteria across the study:
- Stratification (per the ISO 81060-2:2018/Amd-2:2024):
- Subject Group 1:
- At least 3 patients shall be <1000 g in weight
- At least 3 patients shall be 1000 to 2000 g in weight
- At least 3 patients shall be > 2000 g in weight
- At least 3 patients shall be ≥ 29 days and < 1 year of age
- At least 3 patients shall be ≥ 1 year and < 3 years of age
- The remaining patients may be from any of the above age or weight groups in order to complete the sample size
- A patient can be in more than one category simultaneously
- Subject Group 2:
- At least 30% male, 30% female
- Limb circumference distributed as specified in ISO81060-2:2018/Amd2:2024.
- Subject Group 3:
- At least 30% male, 30% female
- Limb circumference distributed as specified in ISO81060- 2:2018/Amd-2:2024.
- At least 10% SBP ≤ 100mmHg
- At least 10% SBP ≥ 160mmHg
- At least 10% DBP ≤ 70mmHg
- At least 10% DBP ≥ 85mmHg
Exclusion Criteria9
- Inability to place the study device appropriately due to patient's anatomy or condition
- Known pregnancy or lactating women (self-report)
- Patients treated with an intra-aortic balloon pump
- Aortic and mitral regurgitation (> 2 nd degree)
- Measurements taken in the lateral position
- Currently participating in another clinical trial with any product, treatment and/or medication that clinically interferes with the study endpoint(s)
- If valid SBP reference measurements for lateral difference is > 15 mmHg (except for neonates with umbilical A-line)
- If valid DBP reference measurements for lateral difference is > 10 mmHg (except for neonates with umbilical A-line)
- At the Principal Investigator's discretion, subject does not qualify to participate in the study for any reasons that are not mentioned above
Interested in this trial?
Get notified about updates and connect with the research team.
Interventions
All subjects enrolled in this study will receive one Philips non-invasive blood pressure cuff. This cuff will be connected via compatible hose to the Philips IntelliVue X3 monitor hosting the released or investigational NIBP algorithm. The invasive blood pressure from the radial arterial line will also be connected to the Philips IntelliVue X3 monitor. The Philips IntelliVue X3 monitor will in turn be connected to a laptop for data collection.
Locations(5)
View Full Details on ClinicalTrials.gov
For the most up-to-date information, visit the official listing.
NCT05673408